Fig. 5

From: TiMON: a real-time integrated monitor for improving the placement and wear of emergency tourniquets

Fig. 5

Blinded and unblinded results from untrained volunteers utilizing the TiMON for emergency tourniquet applications. Red dashed lines indicate the 180 to 300 mmHg range considered for successful tourniquet application. a Applied pressure, where mean comparison p-value = 0.26 and variance comparison p-value < 0.0001. b Application time, where mean comparison p-value = 0.036 and variance comparison p-value = 0.063. c Absolute difference from minimum threshold pressure of 180 mmHg, where mean comparison p-value < 0.0001 and variance comparison p-value < 0.0001